Dual Source CT - Perfusion Defect Visualization with Spiral Dual Energy Scanning

SOMATOM Definition Dual Energy scanning

Author: Benoît Ghaye, MD; Jean-François Monville, MD
Department of Medical Imaging University Hospital of Liège Belgium

History

A 76 year-old male presented with sudden onset of dyspnea to the Emergency Department. Pulmonary embolism was suspected and the patient was referred to our Department for CT pulmonary angiography. A Dual Source CT scan of the thorax was performed on the SOMATOM Definition using spiral Dual Energy evaluation.

 

Diagnosis
After the Dual Energy scan, massive pulmonary


Comments
In the Dual Energy mode two x-ray sources are operated simultaneously at different kV levels of 80 kV and 140 kV. The resulting two spiral data sets acquired in a single scan provide diverse information, which allows differentiating, characterizing, isolating and distinguishing the imaged tissue and material. In the thorax, Dual Energy mode allows investigating lung perfusion after iodinated contrast medium injection. In patients with pulmonary embolism, the Dual Energy mode provides new areas of research, as both morphologic and functional information on the effect of intravascular clots on lung perfusion are obtained using a single modality.
 

Fig 1+2: Excellent visualization of perfusion defects – made possible by Spiral Dual Energy scanning.
 

Examination Protocol

 

Scanner
SOMATOM Definition
Scan area
Thorax
Scan length 342 mm
Scan time 8 s
Scan direction Caudo-cranial
kV
140 kV and 80 kV
Rotation time
0.5 s
Effective mAs 51 eff. mAs / 213 eff. mAs
Slice collimation
0.6 mm
Reconstructed slice thickness
1.5 mm
Spatial Resolution
0.33 mm
Increment 0.5 mm
Kernel
B30f
CTDIvol 6.96 mGy
Contrast Amount
100 ml Enhance / 50 ml NaCl
Contrast Flow rate 5 ml/sec
Start Delay CARE Bolus

 

Date: 2007-09-12


Case Studies

The information presented is for illustration only and is not intended to be relied upon by the reader for instruction as to the practice of medicine. Any health care practitioner reading this information is reminded that they must use their own learning, training and expertise in dealing with their individual patients. This material does not substitute for that duty and is not intended by Siemens Healthcare to be used for any purpose in that regard.

The drugs and doses mentioned herein are consistent with the approval labelling for uses and/or indications of the drug. The treating physician bears the sole responsibility for the diagnosis and treatment of patients, including drugs and doses prescribed in connection with such use. The Operating Instructions must always be strictly followed when operating the system. The source for the technical data is the corresponding data sheets. Results may vary.

Did this information help you?
Thank you for your response
We detected numbers and/or an '@' symbol in your comment. Are you trying to enter a phone number or email address so that we may contact you, please contact us via our email form instead.

Note: to get a direct personal response, please contact us via our email form.

Thank you